Chapter 4 – Functions of the instrument
UM-22100H-U User manual ATEQ D520 Page 94/122
Press on the START key to
start the learning cycle. At the
end of the cycle the display
gives the test result in the
calibration unit.
PRESS = 20.0 mbar
FLOW = 160 XXXXX
TEST
Once this is done, the
calibration cycle stops of its
own accord.
PRESS = 20.0 mbar
FLOW = 160 XXXXX
READY (OK)
Note: the CDF (Leak/Flow calibrator) can be used to calibrate the instrument and
transmit the value via the infra-red link.
3.6.2. CAL check
This special cycle allows the checking of the adjustment (calibration) in calibrated mode.
Refer to the previous paragraph. The CAL check cycle controls if the calibration has not
drifted beyond the limits imposed as a percentage. In the event of an exceeding of the
limits a warning will be set off and a calibration cycle or a checking of the instrument will
be necessary.
Otherwise the "good part” and "cycle end" or "bad part" and "cycle end" indicators will
be activated depending on the measurement made in relation to the reject thresholds.
The CAL check request can be carried out on input 7 of the J3 connector programmed
for this function, or through a programmed function key on the optional RC5 console if it
is installed.
From the main menu, place
the cursor in front of the
CYCLE SPE option and
confirm with ENTER.
RUN PROG.: 001
PARAMETERS
SPE CYCLE: Disabled
Next, place the cursor in front
of CAL check and validate
with the ENTER key.
Infinite fill
CAL.learning
CAL.check
Press on the START key to
start the checking cycle. The
checking cycle stops of its
own accord.
RUN PROG. : 01
PARAMETERS
SPE CYCLE:CAL check
At the end of the cycle, the
display gives the result of the
test in the calibration unit.
PRESS = 20.0 mbar
FLOW = 160 XXXXX
READY (OK)