Chapter 4 – Functions of the instrument
UM-22100H-U User manual ATEQ D520 Page 95/122
3.6.3. Single calibration
The "Single CAL" function or single calibration allows to validate or not the current
calibration for all the programs who measuring in personal calibration mode.
This is allowing carrying out only one calibration learning cycle that will be valid for all
the measurement programs under the personal measurement "CAL" units.
The validation of the single calibration parameter is doing like the following kind:
In the
"PARAMETERS/PrXXX/
FUNCT" menu, place the
cursor in front of "SINGLE
CAL".
PARAM
/
P
r
001
/
F
U
NCT
/
R
.
Drift CAL.: 020 %
NAME :
SINGLE CAL : No
Validate by "Yes" the function
"SINGLE CAL". So the
calibration is valid.
PARAM
/
P
r
001
/
F
U
NCT
/
R
.
Drift CAL.: 020 %
NAME :
SINGLE CAL : Yes
Note: the programs that stay with classical measurement units (no personal calibration)
are not concerned by the single calibration.