Chapter 4 – Functions of the instrument
UM-22100H-U User manual ATEQ D520 Page 61/122
2.1.3. 17) Automatic connector
The automatic connector is a pneumatic command allowing the control of an external
logic (pneumatic seal jig). This command is carried out as soon as the cycle is started
and stopped at the end of the cycle (refer to paragraph 2.1.3.22) "Cycle end").
If several programs are cycled the automatic connectors are activated with the set times
in the first program and are deactivated with the time set for the last program in the
cycle.
They remain active during all the cycles between the first and the last program in the
chain.
The various coupling times are respected by the intermediate programs.
Associated parameters to be adjusted: COUPLING A, COUPLING B.
) Select the option and adjust settings if necessary.
2.1.3. 18) Valve codes
The instrument has 8 electrical programmable outputs (24V DC / 100 mA maximum) of
which 4 may be reserved through the presence of the stamping (valve code 1 internal
and external) and external dump (valve code 2 internal and external). These outputs
(one or more) may be assigned to program numbers. They are generally used to select
the valves in a sequence of cycles. The desired outputs can be activated for each
program (see Chapter 1 paragraph 2.2.1.1.).
Associated parameters to be set: External 1, External 2, External 3, External 4, External
5, External 6, Internal 1, Internal 2.
) Select the option and adjust settings if necessary.