Chapter 4 – Functions of the instrument
UM-22100H-U User manual ATEQ D520 Page 70/122
2.1.3. 26) Display
The "DISPLAY" function allows modify the result display format, this to put the comma
and display the decimal number following the user preferences. Example: X.XXX,
XX.XX, XXX.X or XXXX.
First check or validate the
"Display" function into the
CONFIGURATION menu.
MENU /CONFI/EXTEND
SIGN : No
F. S. DAC : No
DISPLAY : Yes
Then in the test program
parameters, choose the
display format by using the UP
and DOWN keys.
PARAM/Pr001/FUNCTION
STD. COND : No
FILL MODE : No
DISPLAY : X.XXX <
2.1.3. 27) Flow regulation (Option)
The flow regulation function allows to automatically adjusting the pressure following a
flow instruction. This option is available only with an instrument fitted with an electronic
regulator.
It's necessary to deactivate
(No) the "Regulator learn"
function into the
"CONFIGURATION" menu.
MAIN /CONFIGURATION
READY STATUS : No
BARGRAPH : No
REGUL LEARN : No
Validate (Yes) the
"FLOW REG." function in the
"CONFIGURATION/
Extended Menus" menu.
CONFI/EXTENDED
F. S. DAC : No
DISPLAY : No
FLOW REG. : Yes
In the test program
parameters menu, validate
(Yes) the "FLOW REG."
function.
PARAM/Pr001/FUNCTION
FILL MODE : No
DISPLAY : X.XXX
>FLOW REG. : Yes
In the test program
parameters menu, validate
(Yes) the "FILL MODE"
function.
PARAM/Pr001/FUNCTION
>FILL MODE : Yes
DISPLAY : X.XXX
FLOW REG. : Yes
Then select the "PERCENT"
mode in the "FILL MODE"
function.
r001/FUNCT/FILL MODE
>FILL MODE : PERCENT
Connect a pass part to the instrument and make the flow instruction adjustment by
starting the "Elec Reg Adjust" special cycle.