EasyManua.ls Logo

ATEQ D520 - Page 80

ATEQ D520
134 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 4 – Functions of the instrument
UM-22100H-U User manual ATEQ D520 Page 70/122
2.1.3. 26) Display
The "DISPLAY" function allows modify the result display format, this to put the comma
and display the decimal number following the user preferences. Example: X.XXX,
XX.XX, XXX.X or XXXX.
First check or validate the
"Display" function into the
CONFIGURATION menu.
MENU /CONFI/EXTEND
ED
SIGN : No
F. S. DAC : No
DISPLAY : Yes
Then in the test program
parameters, choose the
display format by using the UP
and DOWN keys.
PARAM/Pr001/FUNCTION
STD. COND : No
FILL MODE : No
DISPLAY : X.XXX <
2.1.3. 27) Flow regulation (Option)
The flow regulation function allows to automatically adjusting the pressure following a
flow instruction. This option is available only with an instrument fitted with an electronic
regulator.
It's necessary to deactivate
(No) the "Regulator learn"
function into the
"CONFIGURATION" menu.
MAIN /CONFIGURATION
READY STATUS : No
BARGRAPH : No
REGUL LEARN : No
Validate (Yes) the
"FLOW REG." function in the
"CONFIGURATION/
Extended Menus" menu.
CONFI/EXTENDED
MENUS
F. S. DAC : No
DISPLAY : No
FLOW REG. : Yes
In the test program
parameters menu, validate
(Yes) the "FLOW REG."
function.
PARAM/Pr001/FUNCTION
FILL MODE : No
DISPLAY : X.XXX
>FLOW REG. : Yes
In the test program
parameters menu, validate
(Yes) the "FILL MODE"
function.
PARAM/Pr001/FUNCTION
>FILL MODE : Yes
DISPLAY : X.XXX
FLOW REG. : Yes
Then select the "PERCENT"
mode in the "FILL MODE"
function.
r001/FUNCT/FILL MODE
>FILL MODE : PERCENT
Connect a pass part to the instrument and make the flow instruction adjustment by
starting the "Elec Reg Adjust" special cycle.

Table of Contents