EasyManuals Logo

Beckman Coulter UniCel DxC Synchron 800 User Manual

Beckman Coulter UniCel DxC Synchron 800
510 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #255 background imageLoading...
Page #255 background image
Weekly Maintenance
Replace Cuvette Wiper
UniCel DxC Systems Instructions For Use A13914 Maintenance
October 2005 Page 10-9
10
Replace Cuvette Wiper
With constant use, the cuvette wiper can deteriorate and become stained and torn. The
cuvette wiper should be changed weekly or as needed to maintain the cleanliness and
integrity of the cuvette.
Step Action
1 Remove the cuvette wash station cover. Pull up on the two locking pins (1).
Pivot the right side of the cover, turning it forward. Lift out the cover,
being careful not to disturb the CC sample mixer.
Figure 10.1
2 Remove the two thumbscrews (1) on each side of the cuvette washer. Lift
off the upper section of the cuvette wash assembly. Pull off the silicone
wiper.
Figure 10.2
(1 of 2)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Beckman Coulter UniCel DxC Synchron 800 and is the answer not in the manual?

Beckman Coulter UniCel DxC Synchron 800 Specifications

General IconGeneral
BrandBeckman Coulter
ModelUniCel DxC Synchron 800
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals