Weekly Maintenance
Clean Total Protein and Albumin Cup Modules (DxC 800 Only)
UniCel DxC Systems Instructions For Use A13914 Maintenance
October 2005 Page 10-11
10
Cleaning Preparation
Prepare the following solutions before beginning the cleaning procedure:
Cleaning Procedure
1N HCL solution Combine one part 6N HCL and five parts of deionized
water for cleaning Total Protein cup.
OR
10% Cleaning Solution Combine one part Wash Concentrate II with nine parts
of deionized water for cleaning Albumin cup.
Step Action
1Select:
the Utils icon from the menu bar,
<2> Maintenance,
<4> Cup Maintenance,
<Total Protein> and <Albumin> buttons.
2Select <Rinse>.
The cups are drained and rinsed 5 times with water. A Result window
appears. DO NOT close this window at this time.
3 When rinsing is complete, use a transfer pipette to manually add:
One mL of 1N HCL solution to the Total Protein cup.
One mL of 10% Cleaning solution to the Albumin cup.
4 Let the cleaning solution sit in the cups for 10 minutes. The stir bars will
continue to stir during this time.
5 At the end of the 10 minute waiting period, select <Close> to exit the
Result window.
6Select <Rinse>.
The cups are drained and rinsed 5 times with water.
7 When rinsing is complete select <Close> to exit the Result window then
select <Drain>.
8 When draining is complete select <OK> to continue. When activity has
completed select <Close> to exit the Result window, and select <Cancel>
to return to the Maintenance Menu.
(1 of 2)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Beckman Coulter UniCel DxC Synchron 800 and is the answer not in the manual?

Beckman Coulter UniCel DxC Synchron 800 Specifications

General IconGeneral
BrandBeckman Coulter
ModelUniCel DxC Synchron 800
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals