Monthly Maintenance
Replace Alkaline Buffer
UniCel DxC Systems Instructions For Use A13914 Maintenance
October 2005 Page 10-19
10
5 Remove reagent input line #30 (1) (line draining into damper) from the top
of the in-line filter.
Figure 10.7
6 Twist off the in-line filter (P/N 669212) from the input line and replace.
Note: When installing in-line filters, be sure the arrow on the filter is
pointing in the direction of reagent flow.
7 After installation of the filter is complete, reattach reagent input line #30
(that flows into damper) to the top of the filter.
8 Select the Rgts/Cal icon from the menu bar.
9Select <MC Rgts>.
10 Select Load [F1].
Step Action, continued
(2 of 3)