As-Needed/As-Required Maintenance
Flow Cell Flushing (Manual)
Maintenance UniCel DxC Systems Instructions For Use A13914
Page 10-98 October 2005
4 Remove line #23 from the bottom of the flow cell. Attach the connector
fitting (1) on the small diameter tubing, included in the Flow cell flush kit,
to the syringe. Connect the open end of the tubing to the bottom of the flow
cell.
Figure 10.71
5 Without removing the hemostat, remove line #35 from the top of the flow
cell. Remove line #27 from the front of the flow cell in order to allow
residual acid to leak out.
6 Draw back on the syringe. This will withdraw any remaining solution left
in the flow cell and remove any material that may be causing the occlusion
of the small lumen.
Note: The procedure may be concluded here if the problem has been
resolved. Proceed to Step 9.
If the problem is not resolved, continue to the next step.
7 Re-attach line #27 to the front of the flow cell. Remove the syringe and
empty the contents. Fill the syringe with ISE Buffer.
Step Action, continued
(2 of 3)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Beckman Coulter UniCel DxC Synchron 800 and is the answer not in the manual?

Beckman Coulter UniCel DxC Synchron 800 Specifications

General IconGeneral
BrandBeckman Coulter
ModelUniCel DxC Synchron 800
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals