EasyManuals Logo

Beckman Coulter UniCel DxC Synchron 800 User Manual

Beckman Coulter UniCel DxC Synchron 800
510 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #345 background imageLoading...
Page #345 background image
As-Needed/As-Required Maintenance
Flow Cell Flushing (Manual)
UniCel DxC Systems Instructions For Use A13914 Maintenance
October 2005 Page 10-99
10
8 Remove Line #35 from the top of the flow cell. Attach the connector fitting
(1) of the large diameter tubing, included in the Flow cell flush kit, to the
syringe. Connect the open end of the tubing to the top of the flow cell.
Flush the buffer through.
Figure 10.72
9 Reattach the lines to the flow cell and remove the hemostat.
10 Once all components are in place for the flow cell select <OK> to prime
the flow cell. Select <Close> in the pop-up window to exit the ISE Service
screen.
11 Prime the flow cell 5 times by selecting:
the Utils icon from the menu bar,
<1> Prime,
MC [F4],
ISE All check box.
Type 5 in the Number of primes to repeat field.
Select <Start Prime>.
12
CAUTION
Pinching Hazard. When lowering the ISE module, keep fingers clear.
Lower the ISE module down into cavity and reinstall the ISE module
cover.
13 Calibrate the ISE chemistries.
Step Action, continued
(3 of 3)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Beckman Coulter UniCel DxC Synchron 800 and is the answer not in the manual?

Beckman Coulter UniCel DxC Synchron 800 Specifications

General IconGeneral
BrandBeckman Coulter
ModelUniCel DxC Synchron 800
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals