As-Needed/As-Required Maintenance
Washing All Cuvettes With System Wash
UniCel DxC Systems Instructions For Use A13914 Maintenance
October 2005 Page 10-115
10
Step Action
1 This procedure is accessed by selecting:
• the Utils icon from the menu bar,
• <2> Maintenance,
• <6> Wash All Cuvettes.
Follow all instructions listed on the screen.
2Select <Continue> to perform the procedure.
If desired, you may exit this procedure at any time before it is completed
by selecting:
• STOP on the instrument; wait for system to finish processing,
• <OK> to clear pop-up message,
• <Cancel> to exit the procedure screen,
• the Instr Cmd icon from the menu bar,
• <1> Home.
3Select <Close> at the completion of this procedure to exit the pop-up
Result window.
4 Exit Maintenance menu.
NOTICE
To resume autoprime and return system to Standby, you must exit
Maintenance menu.