Result Errors and Codes
UniCel DxC Systems Instructions For Use A13914 Troubleshooting Calibration and Result Errors
October 2005 Page 13-19
13
DIVISION ERROR ZD Division Error – Special calculation
divisor is zero.
HI REAGENT MC reagent level high.
INCOMPLETE AB Calculation incomplete.
INSUFF DIL1 Insufficient diluent reagent for
scheduled tests.
INSUFF REAG Insufficient reagent strength.
INVALID CHEM Invalid chemistry.
LEV SENSE ERR Low reagent fluid.
LO REAGENT MC reagent level low.
MOTION ERR Motion error.
NO DIL1 ON SYS No diluent reagent on the reagent
carousel.
NO REAG ON SYS Reagent not on board.
NO SAMP DETECT Low sample fluid.
NO SAMP ON SYS Sample not on board.
NOT ACCEPTED Wrong state for sample program
message.
PENDING TEST Not ready.
PROBE OBSTRUCT Probe obstruction detected.
RACK CONFLICT HDLC/IBCT/HbA1c sample in wrong
rack, test cancelled.
RESULT ERROR Result error.
Results Suppressed AG EXCESS AX Antigen excess.
Results Suppressed BL ABS HI BH Blank absorbance high.
Results Suppressed BL ABS LO BL Blank absorbance low.
Results Suppressed BL MAX DEV BO Blank outlier (maximum deviation).
Results Suppressed BL MEAN DEV BN Blank noise (mean deviation).
Results Suppressed BL RATE HI SH Blank rate high.
Results Suppressed BL RATE LO SL Blank rate low.
Results Suppressed CAL REF DRIFT EC Excessive reference drift – sample to
calibration.
Results Suppressed INI ABS HI AH Initial absorbance too high.
Results Suppressed INI ABS LO AL Initial absorbance too low.
Results Suppressed INI RATE HI IR Initial rate too high.
Table 13.9 Result Errors and Codes, continued
Result Displayed on Report Printed in
Remarks Area
Host
Code
Inst.
Code
Description
(2 of 3)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Beckman Coulter UniCel DxC Synchron 800 and is the answer not in the manual?

Beckman Coulter UniCel DxC Synchron 800 Specifications

General IconGeneral
BrandBeckman Coulter
ModelUniCel DxC Synchron 800
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals