Page #440 background image
Result Errors and Codes
Troubleshooting Calibration and Result Errors UniCel DxC Systems Instructions For Use A13914
Page 13-20 October 2005
Results Suppressed INI RATE LO IL CC chemistries, initial rate too low.
MC chemistries, a stirrer failure.
Results Suppressed INIT ADC HI HI Initial ADC error high.
Results Suppressed INIT ADC LO LI Initial ADC error low.
Results Suppressed INIT COND HI CH Initial conductance high.
Results Suppressed INIT COND LO CL Initial conductance low.
Results Suppressed K REQ IK Bad K value.
Results Suppressed NA/K REQ IN Bad NA value.
Results Suppressed OIR HI DH Out of instrument range high.
Results Suppressed OIR LO DL Out of instrument range low.
Results Suppressed ORDAC HI OH Out of instrument range ORDAC high.
Results Suppressed ORDAC LO OL Out of instrument range ORDAC low.
Results Suppressed URDAC HI GH Under range URDAC high.
Results Suppressed URDAC LO GL Under range URDAC low.
Results Suppressed ORR HI UH Out of reportable range high.
Results Suppressed ORR LO UL Out of reportable range low.
Results Suppressed ORR O HI UO Out of ORDAC reportable range high.
Results Suppressed OVERFLOW OF Number overflow error.
Results Suppressed REF NOISE DR Reference signal noise (ISE only).
Results Suppressed RX ABS HI HR Reaction absorbance high.
Results Suppressed RX ABS LO LR Reaction absorbance low.
Results Suppressed RX ERR RE Reaction error.
Results Suppressed RX MAX DEV RO Reaction outlier (maximum deviation).
Results Suppressed RX MEAN DEV RN Reaction noise (mean deviation).
Results Suppressed RX NOISE EA A Erratic ADC (Cup Chems only).
Results Suppressed RXN RATE HI RH Reaction rate high.
Results Suppressed RXN RATE LO RL Reaction rate low.
Results Suppressed SAMP NOISE DS Sample signal noise (ISE only).
Results Suppressed SAMP REF DRIFT ES Excessive reference drift – sample to
sample.
Results Suppressed SUB DEPL SD Substrate depleted.
Results Suppressed TEMP ERR TM T Temperature error (MC only).
TEST STOPPED Test was aborted, system issue.
TOO MANY TESTS Too many tests on CC.
Table 13.9 Result Errors and Codes, continued
Result Displayed on Report Printed in
Remarks Area
Host
Code
Inst.
Code
Description
(3 of 3)

Table of Contents

Questions and Answers: