5.2 Getting Started
40
5.2.2 Re-adjusting the Detector
Proceed as follows to ensure that the detector is in the optimum
position, i.e. exactly vertical above the source:
Loosen holding of detectors slightly, so you can move it.
Push <run> button of evaluation unit. The measured counts are
displayed.
Carefully move the detector forward and back to find the posi-
tion where it is measuring the most counts.
If you have found the correct position, fix the detector with
screws so it cannot move anymore, not even if vibrations occur.
Push <run> once more to stop the measurement.
Rod source
Figure 25: Detector window facing the source
Schematic layout
Detector holder
Detector
Detector window