Section 6 Evaluation Unit LB 442
49
6.2 Software Functions and System Configuration
The factory settings are listed in the Configuration Checklist in the
Appendix. Enter the final operating settings in this list. Please keep
in mind that the entry in all operating levels is made with the
<^^^> key and the new value is accepted with <enter>.
The menu structure on pages 147 to 150 provide a general func-
tions overview.
General Data
Password:
You can enter a number comprising up to 6 digits. Upon confirma-
tion with <enter>, the system is protected against unauthorized
manipulation of the parameters. All parameters can still be viewed,
however. Password protection is canceled when you enter the cor-
rect numerical value and confirm your entry with <enter>.
Date:
Enter the current date in the format TT.MM.YY. The correct date is
important for automatic correction of the activity decay of the rod
source.
Decay compensation is performed every night at midnight or when
pressing the <run> button to start the measurement modes, when
the last decay compensation took place more than 24 hours ago.
The tare count rate and the count rates of the individual calibration
points are corrected.
Time:
Enter the current time in the format HH.MM.
System/Version:
Display of instrument type and software version.
Detector software version:
Display of the software version of the connected detector.
Language:
You can choose German, English or French as dialog language by
pressing the <^^^> key.