11.4 Evaluation Unit LB 442
94
11.4.3 Error Code Listing with Signal Output
No.
Error text
In case of error,
measurement:
Alarm
Relay
0/4 -
20mA
ABORT CONTINUE
1
Pulse rate overflow abort continue alarm value
2
No pulses from detector stop continue alarm value
3
HV wrong abort continue alarm value
5
Measuring range for current
output 0/4-20mA faulty
abort continue no alarm normal
6
Detector temperature > 65°C continue continue alarm normal
7
Trouble with detector
communication
abort abort alarm value
9
Detector temperature > 85°C abort abort alarm value
10
No calibration abort abort no alarm hold
11
Power fail
>1 month – time?
abort continue alarm value
12
No input allowed continue continue no alarm normal
32
Wrong password continue continue no alarm normal
34
Wrong input value continue continue no alarm normal
35
Calibration error abort abort no alarm hold
38
Wrong product abort abort no alarm normal
39
Time constant 1 < 1s continue continue no alarm normal
40
Time constant 2 < 1s continue continue no alarm normal
41
min. measuring range not
reached
abort abort no alarm normal
42
max. measuring range ex-
ceeded
abort abort no alarm normal
43
Rate Overflow continue continue no alarm normal
44
Current input error abort continue alarm normal
46
Check external mass counter
setting
abort abort alarm value
47
Check external mass counter
setting
continue continue alarm value