56926BA2, Rev. 02, 04/2019
Avg. HV
limit
exceeds the
Default-HV by
more than 40%.
S
Check Default HV for plausibility and
change if necessary.
Check detector electronics and
photomultiplier.
If necessary, contact BERTHOLD.
ure
sensor
sensor of the
detector is
S
The hardware is defective and, if
necessary, must be checked and
ure
warning
detector
temperature is
close to the
upper or lower
limit of the
permissible
operating
S
If the detector is too hot: Install/check
water cooling system.
If the detector is too cold: Install/check
heating.
ure alarm
detector
temperature has
exceeded the
upper limit of the
permissible
operating
temperature or
dropped below
F
Proper function of the device can no
longer be guaranteed.
It is recommended to have the device
checked by Berthold Technologies, even
if it still seems to work properly.
malfuncti
on
state machine of
the detector has
F
If the event occurs frequently, contact
BERTHOLD.
Recording
plateau recording
C
No action is required. The detector
returns automatically to measurement
mode after plateau recording is finished.
high
reached the
upper HV limit
(HV > 1287V,
bzw. HV > 1300V
F
Detector must be checked and replaced
if necessary.
low
reached the
lower HV limit
(HV < 303V, bzw.
F
Detector must be checked and replaced
if necessary.