56926BA2, Rev. 02, 04/2019
(Meas)
the measuring
F
The photomultiplier may be defective.
Check the photomultiplier or contact
(Ctrl)
the control
S
The photomultiplier may be defective.
Check the photomultiplier or contact
(Aux)
the auxiliary
S
The photomultiplier may be defective.
Check the photomultiplier or contact
averaging
fluctuates more
than permissible.
F
If the event occurs frequently, contact
BERTHOLD.
(Meas)
the measuring
channel deviates
from the
admissible set
F
If the event occurs frequently, contact
BERTHOLD.
(Ctrl)
the control
channel deviates
from the
admissible set
S
If the event occurs frequently, contact
BERTHOLD.
(Aux)
the auxiliary
channel deviates
more than
allowed from the
S
If the event occurs frequently, contact
BERTHOLD.
(RID)
the RID channel
deviates from the
admissible set
S
If the event occurs frequently, contact
BERTHOLD.
voltage -
PMT
high-voltage
supply of the
photomultiplier
F
Check detector electronics and
photomultiplier.
If necessary, contact BERTHOLD.
limited
voltage deviates
from the average
high voltage by
F
Check detector electronics and
photomultiplier.
If necessary, contact BERTHOLD.
Avg. HV
limit
more than 20%
lower than the
Default-HV.
F
Check Default HV for plausibility and
change if necessary.
Check detector electronics and
photomultiplier.
If necessary, contact BERTHOLD.