Appendix A Maintaining the Imaging System
2. Insert the tip of a screwdriver as far as it will go under the center of the fuse drawer.
3. Tug hard on the screwdriver to open the fuse drawer.
4. Remove the drawer from the power module.
110 | ChemiDoc Touch Imaging System