Chapter 3 Acquiring an Image
Workflow for Acquiring a Chemiluminescent Image
Note: For information on acquiring images other than chemiluminescent blot images,
see Workflow for Acquiring Gel and Blot Images on page 62.
The basic steps to acquire a chemiluminescent image are
n Specify the image size.
n Choose the application type.
n Specify the region of interest.
n Set the exposure time.
n Acquire the image.
Note: You can specify the acquisition settings at any point in the workflow,
for example, before you prepare the sample tray or after the tray is in the
transilluminator drawer. The imager retains the settings until you change them.
Specifying the Image Size
You can select a preset image size or specify a custom size.
To specify the image size for a chemiluminescent image
1. In Live View, tap IMAGE SIZE.
2. To choose a standard size, tap Small, Medium, or Large.
46 | ChemiDoc Touch Imaging System