Environmental Requirements
Autofocus Precalibrated focus for any zoom setting
Image flat fielding Dynamic; precalibrated and optimized per application
Autoexposure 2 user-defined modes (rapid or optimal) for chemiluminescent
applications
2 user-defined modes (intense or faint bands) for fluorescent and
visible applications
Environmental Requirements
The ChemiDoc Touch imager requires a space 52 x 56 x 62 cm (W x H x D) and a
clearance of at least 8 cm from the back for instrument ventilation and for connecting or
disconnecting the AC power cord. Place the imager on a sturdy and level laboratory bench
or table away from excessive heat and moisture. The imager’s operating temperature
range is 10–2C. The imager contains a universal power supply that supports a voltage
range of 100240 VAC.
WARNING! Transilluminators are powerful sources of UV radiation, which
can cause serious damage to unprotected eyes and skin. The accessory UV
shield (cat. #1708375) provides UV protection. However, this shield does not
guarantee complete protection nor does it provide protection to others in the
area around the imager. Before performing band excision, the user and any-
one near the imager must put on protective gear including eyeglasses (labor-
atory glasses provide adequate protection), a face shield, lab coat, and
gloves.
User Guide | 23