About Exposure Times for Chemiluminescent Images
resolution; however, the imaging time might increase. Selecting a setting to the right of the
default setting might reduce the resolution below that required for publication; however, the
exposure time is reduced.
Bio-Rad recommends that you start with the optimal binning setting, examine the result,
and then adjust subsequent images.
The Image Resolution/Sensitivity Scale
The Image Resolution/Sensitivity scale has five pixel binning settings: 2x2, 3x3, 4x4, 6x6,
and 8x8.
The following icons indicate binning settings.
Icon Description
Binning level
Selected binning level
Optimal setting for publication-quality images
User Guide | 49