44
Clinician's Guide
Electrode Base Set
The electrode base set is used to customize the electrode positions for home-use Orthoses.
Electrode Base Screw and Washer Set
The electrode base screw and washer set is used to attach electrode bases #1, #2, and #4
to home-use H200 Wireless Orthoses. Use one screw and one washer for each electrode
base.
H200 Wireless Cloth Electrodes
The H200 Wireless Orthosis uses five cloth electrodes. The electrodes insert to the electrode
bases. The cloth electrodes are made from a non-woven cotton/polymer fabric designed to
retain moisture while the H200 Wireless Orthosis is on the arm. The cloth electrodes must
be wet before use and after every three to four hours of use.
The thenar cloth electrode is available in regular and large sizes.
CAUTION: Do not use the NESS H200 Wireless System without the cloth
electrodes.
CAUTION: The H200 Wireless cloth electrodes must be wet before use and
wet again after every four hours of use.
CAUTION: The H200 Wireless cloth electrodes are for single patient use.
CAUTION: The H200 Wireless cloth electrodes must be replaced every two
weeks, or sooner if they become damaged.