81
Chapter 8 - NESS H200 Wireless Software
Figure 8-25: Program C window.
Program CGrasp
Program C activates the flexor muscles only. Pressing Start starts the program. Cyclical
hand flexion is followed by a period of relaxation, and then hand flexion repeats. The program
shuts down automatically after the total program time has elapsed (range: 5–120 minutes).
Stimulation intensity, the duration of flexion and relaxation, and total time are adjustable. See
Figure 8-25.
To configure Program C:
1. Select Program C.
2. Adjust grasp time, stimulation off time, and total time.
3. Press Start. The program countdown clock will appear at the bottom of the screen.
4. If needed, adjust stimulation intensity.
5. Allow the program to continue for the total time or press Stop.
6. Press Done to return to the Program Settings window.
Note: The active program phase will be highlighted in ORANGE.
Note: Adjustments to stimulation intensity affect all programs except Key Open.

Table of Contents