85
Chapter 8 - NESS H200 Wireless Software
Program GMotor Neuromodulation
Program G delivers pulsed stimulation above the motor threshold for an adjustable program
time (5–30 minutes). Pressing Start starts the program. See Figure 8-29.
Figure 8-29: Program G, extensors and flexors window.
To configure Program G:
1. Select Program G.
2. Select Ext. and Flex., Extensors, or Flexors.
Ext. and Flex. (Extensors and Flexors)
1. Adjust extensors on time, stimulation off time, flexors on time, and total time. See Figure
8-29.
2. Press Start. The program countdown clock will appear at the bottom of the screen.
3. If needed, adjust stimulation intensity.
4. Allow the program to continue for the total time, or press Stop.
5. Press Done to return to the Program Settings window.
Note: The active program phase will be highlighted in ORANGE.
Note: Adjustments to stimulation intensity affect all programs except Key Open.