Basic Procedures
H171804E_14_001 39 / 86
5.2 Inserting the Sample plus Spinner into the Magnet
The raising  and lowering of  the sample  is controlled by  a stream of  pressurized air.  Before
placing a sample onto the top of the bore, make sure that the air flow is present (it is quite
audible) and remove any sample which may be rising to the top of the bore.
• On the menu bar, click Acquire.
• On the Sample button, click the drop-down arrow to see more options.
• In the list, select Eject sample manually (ej). The sample lift is turned on.
Wait until the sample lift air is turned on and remove the sample which may be in the
magnet.
• Place the sample plus the spinner on top of the magnet bore.
• On the Sample button, click the drop-down arrow to see more options.
• In the list, select Insert sample manually (ij).
5.3 Locking the Sample
Deuterated solvents are used to generate the signal to be detected and monitored by the lock
system. The frequency and strength of this signal will depend on the solvent used. The main
feature of the TopSpin lock routine is that it sets parameters such as the lock power, gain and
frequency  to a  value  appropriate  to  the  solvent.  With  these default  values  set close  to  that
which would be  expected for that solvent,  the BSMS  can quickly locate  and  "lock“ onto the
solvent  signal  by  sweeping  through  a  range  of  frequencies  or  magnetic  field  values.  The
solvent dependent parameters are taken from the "edlock“ table.
• On the Workflow button bar, click Lock.
• In the Solvents table list, select DMSO and click OK.
Figure5.2: Solvents Table