Basic Procedures
40 / 86 H171804E_14_001
Note: If lock has been achieved, the signal should be of the form of a horizontal line with
some associated noise or ripples (see figure below). The height of this line is called the lock
level.
Figure5.3: Lock Display after Locking the Sample
5.4 Tuning and Matching the Probe
The  sensitivity  of  any  probe  will  vary  with  the  frequency  of  the  signal  transmitted  to  it  and
there  exists  a  frequency  at  which  the  probe  is  most  sensitive.  Furthermore  this  frequency
may  be  adjusted  over  a  certain  range  using  tuning  capacitors  built  into  the  probe  circuitry.
Tuning  involves  adjusting  the  probe  circuitry  so  that  the  frequency  at  which  it  is  most
sensitive is the relevant transmission frequency (SFO1, SFO2 etc.) Each coil in the probe will
be  tuned  (and  matched)  separately.  If  the  probe  has  been  changed  or  the  transmission
frequency altered significantly, it may be necessary to retune the probe. For routine work in
organic solvents with selective probes, the value of the transmitted frequencies are unlikely to
vary greatly. Hence, once the probe has been initially tuned, slight variations in frequency will
not warrant retuning. Typically the transmitted frequency would need to be altered by at least
100kHz to warrant  retuning. However for broadband probes the frequencies transmitted will
vary greatly  from nucleus to  nucleus and  so the probe  will need to  be tuned  each time the
selected nucleus is altered. Whenever a probe is tuned it should also be matched. Matching
involves  ensuring  that  the  maximum  amount  of  the  power  arriving  at  the  probe  base  is
transmitted  up  to  the  coil  which  lies  towards  the  top  of  the  probe.  This  ensures  that  the
minimum  amount  of  the  power  arriving  at  the  probe  base  is  reflected  back  towards  the
amplifiers (and consequently wasted).
Note: Bruker offers two different types of tuning and matching adjustments. In addition to the
manual adjustments of the tuning and matching capacitors, the probes can be equipped with
an Automatic Tuning Module (ATM). Follow the steps below for either option.
5.4.1 Probes Equipped with ATM using the Automated Tuning Routine
• On the Workflow button bar, click Tune.
The display will switch automatically to the acquisition window and display the wobble curve.
The  tuning  and  matching  is  performed  automatically.  If  multiple  frequencies  are  used  in  a
parameter set such as C13CPD etc., ATMA will start adjusting the lowest frequency first and
will switch in the order of increasing frequency automatically.