Basic Procedures
44 / 86 H171804E_14_001
5.6 Shimming
Shimming is a process in which minor adjustments are made to the magnetic field until the 
field  homogeneity  (uniformity) is  optimized.  Improving  the  homogeneity  will  result  in  better 
spectral resolution. It will be necessary to re-shim each time a probe or sample is changed.
The  system  manager  has  stored  appropriate  shim  values  (in  so  called  shim  files)  for  each
probe that will greatly reduce the shimming time required whenever a probe is changed.
5.6.1 Routine Shimming Using TopShim
This is routine shimming and should be carried out at the beginning of every NMR session,
and whenever the sample in the magnet is changed. Routine shimming involves making fine
adjustments to the Z, Z
2
, Z
3
, Z
4 
and Z
5
 shims. Some higher field magnets may require higher
order  Z  shims.  The  system  administrator  has  programed  TopShim  to  achieve  the  best
homogeneity on each sample and it is fully automatically.
The core method of TopShim is gradient shimming. A quality criterion for the final line-shape
ensures best results for all situations.
TopShim is using for all deuterated solvents the 
2
H gradient shimming method and for other
solvents especially H
2
O, the 
1
H gradient shimming method.
• On the Workflow button bar, click Shim.
The shimming will start momentarily and should take less than a minute.
5.7 Setting the Probe/Solvent Dependent Parameters
Parameters, such as the 90
0
 or the decoupler pulse length with associated power levels can
be stored for any specific nucleus on any available channel. These values have been entered
by the installation engineer or by the system administrator and are automatically loaded using
the Prosol button.
• On the Workflow button bar, click Prosol.
5.8 Adjusting the Receiver Gain
The receiver  gain is  a very  important parameter  that is  used to  match the  amplitude  of the
FID to the dynamic range of the digitizer. The gain is automatically set using the Gain button.
• On the Workflow button bar, click Gain.
The receiver gain adjustment will start momentarily and it should take less than a minute.
5.9 Starting the Acquisition
The  process  of  receiving  the  NMR  signals  is  called  an  acquisition.  Data  is  said  to  be
acquired. When an acquisition is carried out, “raw” data is acquired and the received signal is
called  an  FID  (Free  Induction  Decay).  The  digital  point  size  of  the  FID  is  called  TD,  Time
Domain.
• On the Workflow button bar, click Run.
As a default from the selected experiment, the spectrometer performs a number of dummy-
scans before starting the actual number of scans.