The NMR Sample
52 / 86 H171804E_14_001
7.2 Sample Tube
When  the  sample  is  being  analyzed  it  may  be  rotated,  depending  on  the  type  of  probe  or
experiment. Spinning the sample has the effect of cancelling out field inhomogeneities in the
X  and  Y  direction  and  consequently  improves  the  spectral  resolution.  A  disadvantage  of
spinning  is  that  it  may  lead  to  the  presence  of  spinning  sidebands.  These  are  spurious
signals  (i.e.  peaks)  that  result  from  the  modulation  of  the  magnetic  field  at  the  spinning
frequency. The peaks always appear on either side of any large genuine peak at a separation
equal  to  the  spinning  rate.  The  intensity  of  these  sidebands  will  be  proportional  to  the
intensity of  the genuine peak.  Thus if the  spin rate is  20 revolutions/second (=  20 Hz), you
would  look  for  spinning  side  bands  at  frequencies  20  Hz  above  and  below  the  resonance
frequencies of genuine signals.
Figure7.1: Spectrum Showing Spinning Sidebands
1. 13C Satellites 2. Spinning Sidebands
While the presence of spinning sidebands may be inevitable, their size often depends on the
sample  tube  quality.  Ideally,  the  sample  tube  should  have  perfect  cylindrical  symmetry.
Unusually large sidebands  might suggest that the  tube symmetry is inadequate  and might
warrant using tubes with higher specifications (and of course greater cost).
Sample tubes must always be kept clean and free from dust and scratches. Do not scrub the
tubes with test tube brushes. Be aware that new NMR tubes cannot be assumed to be clean.
The tubes may be cleaned by washing in Acetone or distilled water. Liquid detergent may be
used as long as it is rinsed out within a few minutes to prevent etching of the tube. The tubes
may also be cleaned ultrasonically in an appropriate solution. If all the above measures fail,
the tubes should be soaked in AQUA REGIA for up to two days and then rinsed thoroughly
before being dried. NMR tubes can be oven dried but should not be heated above 100°C as
they can become distorted and subsequently fail to spin as required. Drying is best achieved
by passing filtered nitrogen through the tube.