Basic Procedures
42 / 86 H171804E_14_001
• In the ATMM control window, click the Tuning buttons to move and display the wobble
curve centered.
• In the ATMM control window, click the Matching buttons to adjust the dip of the wobble
curve to the lowest position.
It is very convenient to look at the complex wobble display. If the circle with the red dot is too
big  or  too  small,  then  adjust  matching  first.  Once  the  circle  runs  through  the  center  of  the
coordinate system, bring the red dot into the center by adjusting the tuning. In essence, the
curve tells you in which order to adjust tuning and matching.
Note: The T/M settings will be saved with a click on the Start button.
Since the Tuning and Matching adjustment interact with each other, a repeat of all steps are
necessary for a perfect tune and match. If multiple frequencies are used in a parameter set
such as C13CPD, use the Nucleus Selection radio buttons in the ATMM control window to
switch to another nucleus and repeat the tuning and matching.
The  wobble  routine  works  by  transmitting  a  weak  signal  to  the  probe  and  comparing  the
probe  and  cable  impedance  with  a  50  Ohm  reference  within  the  HPPR.  The  transmitted
frequency is centered on SFO1, SFO2 etc. but "swept“ over a range determined by the size
of the parameter WBSW (see below). The resultant curve is the familiar response curve of a
resonance  circuit  and  is simply  a  measure  of  the  amplitude  of  the  reflected  signal  (vertical
axis) against frequency (horizontal axis).
Matching    involves adjusting  the  probe  so  that  the  minimum  of  the  wobble  curve is  at  the
base  of  the  display  (i.e.  touching  the  horizontal  frequency  axis).  This  represents  minimum
reflection of the transmitted signal.
Tuning involves  ensuring that  this occurs  at the  frequency of  transmission which  will be  at
the center of the screen’s horizontal scale. It will be seen however that the tune and match
adjustments  interact  with  each  other  and  must  be  adjusted  in  tandem.  When  the  wobble
curve minimum is centered and at the base of the screen the probe has been optimally tuned
and matched.