Programmable DC Electronic Load 63600 Series Operation & Programming Manual
:USER
:ADVance
:WAVeform
SP
<NRf+>
?
SP
MAX
MIN
Suffix
<NR1>
<NR1>
<NR1>
:NSELect
:DATA
:POINt
:STATus
:SAVe
:CLEar
:REMain
:EXEcute
:STATus
SP
<arg1>,<arg2>,<arg3>,<arg4>,<arg5>
?
<NR1>
SP
MAX
MIN
<AARD>
<AARD>
<AARD>
SP
<DLABRD>
?
<NR1>
?
?
?
?
?
<NR1>
<NR1>
<NR1>
<NR1>
<NR1>
<NR1>
<NR1>
<NR1>
ADVance:SINE:FREQuency
Type: Channel-Specific
Description: Set frequency for sine wave dynamic mode.
Setting Syntax: ADVance:SINE:FREQuency<space><NRf+>[suffix]
Setting Parameters:<NRf+>, 0.01Hz ~ 20000.00Hz, Resolution = 0.01Hz, Unit = Hertz
Setting Example: ADV:SINE:FREQ 1000 Set frequency = 1kHz.
ADV:SINE:FREQ 1kHz Set frequency = 1kHz.
ADV:SINE:FREQ MAX Set frequency = maximum value.
ADV:SINE:FREQ MIN Set frequency = minimum value.
Query Syntax: ADVance:SINE:FREQuency?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = Hertz]
Query Example: ADV:SINE:FREQ?
ADV:SINE:FREQ? MAX
ADV:SINE:FREQ? MIN
ADVance:SINE:IAC
Type: Channel-Specific
Description: Set AC current for sine wave dynamic mode.
Setting Syntax: ADVance:SINE:IAC<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example: ADV:SINE:IAC 0.5 Set AC current = 0.5A.
ADV:SINE:IAC 500mA Set AC current = 0.5A.
ADV:SINE:IAC MAX Set AC current = maximum value.
ADV:SINE:IAC MIN Set AC current = minimum value.
Query Syntax: ADVance:SINE:IAC?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = Ampere]
Query Example: ADV:SINE:IAC?