Remote Operation
5-29
ADV:SINE:IAC? MAX
ADV:SINE:IAC? MIN
ADVance:SINE:IDC
Type: Channel-Specific
Description: Set DC current for sine wave dynamic mode.
Setting Syntax: ADVance:SINE:IDC<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example: ADV:SINE:IDC 0.5 Set DC current = 0.5A.
ADV:SINE:IDC 500mA Set DC current = 0.5A.
ADV:SINE:IDC MAX Set DC current = maximum value.
ADV:SINE:IDC MIN Set DC current = minimum value.
Query Syntax: ADVance:SINE:IDC?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = Ampere]
Query Example: ADV:SINE:IDC?
ADV:SINE:IDC? MAX
ADV:SINE:IDC? MIN
ADVance:OCP:RESult?
Type: Channel-Specific
Description: Returns the result of OCP test function.
Setting Syntax: None
Setting Parameters:None
Setting Example: None
Query Syntax: ADVance:OCP:RESult?
Return Parameters:<arg1>,<arg2>,<arg3>
<arg1>: Pass/Fail. <NR1>, 0: PASS 1: FAIL [Unit = None]
<arg2>: OCP current. <NR2>, [Unit = Ampere]
<arg3>: Maximum power. <NR2>, [Unit = Watt]
When the returns are
-1,-1,-1 denotes OCP test is stop.
-2,-2,-2 denotes OCP test is ready to execute what wait for Von or
other condition.
-3,-3,-3 denotes OCP test is execute.
Query Example: ADV:OCP:RES?
ADVance:OCP:DWELl
Type: Channel-Specific
Description: Set dwell time for OCP test mode.
Setting Syntax: ADVance:OCP:DWELl<space><NRf+>[suffix]
Setting Parameters:<NRf+>, 10μs ~ 1s Resolution = 10μs, Unit = Second
Setting Example: ADV:OCP:DWEL 0.5 Set off time = 0.5s.
ADV:OCP:DWEL 500msSet off time = 0.5s.
ADV:OCP:DWEL MAX Set off time = maximum value.
ADV:OCP:DWEL MIN Set off time = minimum value.
Query Syntax: ADVance:OCP:DWELl?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = Second]
Query Example: ADV:OCP:DWEL?
ADV:OCP:DWEL? MAX
ADV:OCP:DWEL? MIN