Programmable DC Electronic Load 63600 Series Operation & Programming Manual
5-30
ADVance:OCP:IEND
Type: Channel-Specific
Description: Set end current for OCP test mode.
Setting Syntax: ADVance:OCP:IEND<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example: ADV:OCP:IEND 0.5 Set end current = 0.5A.
ADV:OCP:IEND 500mA Set end current = 0.5A.
ADV:OCP:IEND MAX Set end current = maximum value.
ADV:OCP:IEND MIN Set end current = minimum value.
Query Syntax: ADVance:OCP:IEND?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = Ampere]
Query Example: ADV:OCP:IEND?
ADV:OCP:IEND? MAX
ADV:OCP:IEND? MIN
ADVance:OCP:ISTArt
Type: Channel-Specific
Description: Set starts current for OCP test mode.
Setting Syntax: ADVance:OCP:ISTArt<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example: ADV:OCP:ISTA 0.5 Set starts current = 0.5A.
ADV:OCP:ISTA 500mA Set starts current = 0.5A.
ADV:OCP:ISTA MAX Set starts current = maximum value.
ADV:OCP:ISTA MIN Set starts current = minimum value.
Query Syntax: ADVance:OCP:ISTArt?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = Ampere]
Query Example: ADV:OCP:ISTA?
ADV:OCP:ISTA? MAX
ADV:OCP:ISTA? MIN
ADVance:OCP:SPECification:H
Type: Channel-Specific
Description: Set high level current of specification for OCP test mode.
Setting Syntax: ADVance:OCP:SPECification:H<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example: ADV:OCP:SPEC:H 0.5 Set high level current = 0.5A.
ADV:OCP:SPEC:H 500mA Set high level current = 0.5A.
ADV:OCP:SPEC:H MAX Set high level current = maximum
value.
ADV:OCP:SPEC:H MIN Set high level current = minimum
value.
Query Syntax: ADVance:OCP:SPECification:H?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = Ampere]
Query Example: ADV:OCP:SPEC:H?
ADV:OCP:SPEC:H? MAX
ADV:OCP:SPEC:H? MIN
ADVance:OCP:SPECification:L
Type: Channel-Specific
Description: Set low level current of specification for OCP test mode.
Setting Syntax: ADVance:OCP:SPECification:L<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example: ADV:OCP:SPEC:L 0.5 Set low level current = 0.5A.