Remote Operation
5-71
Setting Parameters:Refer to respective specification for valid value range.
Setting Example: POW:STAT:L2 20 Set the load parameter L2 = 20W.
POW:STAT:L2 10W Set the load parameter L2 = 10W.
POW:STAT:L2 MAX Set the load parameter L2 = maximum
value.
POW:STAT:L2 MIN Set the load parameter L2 = minimum
value.
Query Syntax: POWer:STATic:L2?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = Watt]
Query Example: POW:STAT:L2?
POW:STAT:L2? MAX
POW:STAT:L2? MIN
POWer:STATic:RISE
Type: Channel-Specific
Description: Set the rising slew rate of current for constant power mode.
Setting Syntax: POWer:STATic:RISE<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example: POW:STAT:RISE 2.5 Set rising slew rate to 2.5A/μs.
POW:STAT:RISE 1A/μs Set rising slew rate to 1A/μs .
POW:STAT:RISE MAX Set rising slew rate to the
maximum value of load.
POW:STAT:RISE MIN Set rising slew rate to the
minimum value of load.
Query Syntax: POWer:STATic:RISE?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = A/μs]
Query Example: POW:STAT:RISE?
POW:STAT:RISE? MAX
POW:STAT:RISE? MIN