Programmable DC Electronic Load 63600 Series Operation & Programming Manual
range.
Set the low-level of power specific in sequence:
Arg15: <NRf>, Refer to respective specification for valid value
range.
Set the delay time of Pass/Fail in sequence:
Arg16: <NRf>, 0s ~ 30s, Resolution = 0.0001s, Unit = Second.
Setting Example: PROG:DATA:LIST 1,1,AUTO,CC,2,3.5,0.5,0.5,2,-1,-1,-1,-1,-1,-1,1
Query Syntax 1: PROGram:DATA:LIST?<space><Arg1>,<Arg2>[<space><MAX |
MIN>]
Selects a program:
Arg1: <NR1>, 1 ~ 10, Resolution = 1, Unit = None.
Selects a sequence:
Arg2: <NR1>, 1 ~ N, Resolution = 1, Unit = None.
Query Syntax 2: PROGram:DATA:LIST?<space><Arg1>,<Arg2>,<Arg3>,<Arg4><s
pace><MAX | MIN>
Selects a program:
Arg1: <NR1>, 1 ~ 10, Resolution = 1, Unit = None.
Selects a sequence:
Arg2: <NR1>, 1 ~ N, Resolution = 1, Unit = None.
Selects a run mode:
Arg3: <NRf>, CC(0), CR(1), CV(2), CP(3), Unit = None.
Selects the mode’s range:
Arg4: <NRf>, LOW(0), MIDDLE(1), HIGH(2), Unit = None.
Return Parameters: <aard>
Query Example: PROG:DATA:LIST? 2,1
PROG:DATA:LIST? 2,1 MAX
PROG:DATA:LIST? 2,1 MIN
PROG:DATA:LIST? 2,1,1,0 MAX
PROG:DATA:LIST? 2,1,1,0 MIN
Return Example: 2,1,AUTO,CC,HIGH,3.5,0.5,0.5,2,-1,-1,-1,-1,-1,-1,1
PROGram:DATA:STEP
Type: Channel-Specific
Description: Set the step parameters in program. (Note: All setting parameters in
this command can’t use suffix.)
Setting Syntax: PROGram:DATA:STEP<space><Arg1>,<Arg2>,<Arg3>,<Arg4>,
<Arg5>,<Arg6>,<Arg7>,<Arg8>,<Arg9>,<Arg10>,<Arg11>,
<Arg12>,<Arg13>,<Arg14>,<Arg15>,<Arg16>
Setting Parameters:
Selects a program to be set:
Arg1: <NR1>, 1 ~ 10, Resolution = 1, Unit = None.
Set the trigger mode of sequence:
Arg2: <NRf>, SKIP(0), AUTO(1), MANUAL(2), EXTERNAL(3),
Unit = None.
Set the run mode of sequence:
Arg3: <NRf>, CC(0), CR(1), CV(2), CP(3), Unit = None.
Set the mode’s range of sequence:
Arg4: <NRf>, LOW(0), MIDDLE(1), HIGH(2), Unit = None.
Set the start value according to run mode in sequence:
Arg5: <NRf>, Refer to respective specification for valid value
range.
Set the end value according to run mode in sequence:
Arg6: <NRf>, Refer to respective specification for valid value
range.