Remote Operation
5-75
Set the falling of slew rate in sequence:
Arg7: <NRf>, Refer to respective specification for valid value
range.
Set the rising of slew rate in sequence:
Arg8: <NRf>, Refer to respective specification for valid value
range.
Set the dwell time of sequence:
Arg9: <NRf>, 0.1ms ~ 30s, Resolution = 0.0001s, Unit =
Second.
Set the high-level of voltage specific in sequence:
Arg10: <NRf>, Refer to respective specification for valid value
range.
Set the low-level of voltage specific in sequence:
Arg11: <NRf>, Refer to respective specification for valid value
range.
Set the high-level of current specific in sequence:
Arg12: <NRf>, Refer to respective specification for valid value
range.
Set the low-level of current specific in sequence:
Arg13: <NRf>, Refer to respective specification for valid value
range.
Set the high-level of power specific in sequence:
Arg14: <NRf>, Refer to respective specification for valid value
range.
Set the low-level of power specific in sequence:
Arg15: <NRf>, Refer to respective specification for valid value
range.
Set the delay time of Pass/Fail in sequence:
Arg16: <NRf>, 0s ~ 30s, Resolution = 0.0001s, Unit = Second.
Setting Example: PROG:DATA:STEP ,AUTO,CC,2,3.5,20.0,0.5,0.5,2,-1,-1,-1,-1,-1,-
1,1
Query Syntax 1: PROGram:DATA:STEP?<space><Arg1>[<space><MAX | MIN>]
Selects a program:
Arg1: <NR1>, 1 ~ 10, Resolution = 1, Unit = None.
Query Syntax 2: PROGram:DATA:STEP?<space><Arg1>,<Arg2>,
<Arg3><space><MAX | MIN>
Selects a program:
Arg1: <NR1>, 1 ~ 10, Resolution = 1, Unit = None.
Selects a run mode:
Arg2: <NRf>, CC(0), CR(1), CV(2), CP(3), Unit = None.
Selects the mode’s range:
Arg3: <NRf>, LOW(0), MIDDLE(1), HIGH(2), Unit = None.
Return Parameters:<aard>
Query Example: PROG:DATA:STEP? 1
PROG:DATA:STEP? 1 MAX
PROG:DATA:STEP? 1 MIN
PROG:DATA:STEP? 1,0,2 MAX
PROG:DATA:STEP? 1,0,2 MIN
Return Example: 1,AUTO,CC,HIGH,3.5,20.0,0.5,0.5,2,-1,-1,-1,-1,-1,-1,1
PROGram:NSELect
Type: Channel-Specific
Description: Selects the program number which to be executed.
Setting Syntax: PROGram:NSELect<space><NRf+>