Programmable DC Electronic Load 63600 Series Operation & Programming Manual
channels in one frame. The frame will ignore the setting if the
channel does not exist.
Setting Syntax: CDR<space>“n,n,n,n,n,n,n,n,n,n”
Setting Parameters: <NR2>, [Unit=A/μs]
Example: CDR “1.0,2.5,2.5,10,2.0,5.0,5.0,5.0,0,0”
CDF
Type: All Channels
Description: Set the falling slew rate of CCDL/CCDM/CCDH mode to the ten
channels in one frame. The frame will ignore the setting if the
channel does not exist.
Setting Syntax: CDF<space>“n,n,n,n,n,n,n,n,n,n”
Setting Parameters: <NR2>, [Unit=A/us]
Example: CDF “1.0,2.5,2.5,10,2.0,5.0,5.0,5.0,0,0”
CDRT
Type: All Channels
Description: Set the repeat count of CCDL/CCDM/CCDH mode to the ten
channels in one frame. The frame will ignore the setting if the
channel does not exist.
Setting Syntax: CDRT<space>“n,n,n,n,n,n,n,n,n,n”
Setting Parameters: <NR1>, 0 ~ 65535
Example: CDRT “1,2,2,10,2,5,5,5,0,0”
L
Type: All Channels
Description: Set the load level according to mode setting for the ten channels in
one frame. The frame will ignore the setting if the channel does not
exist.
Setting Syntax: L<space>“n,n,n,n,n,n,n,n,n,n”
Setting Parameters: <NR2> [Unit=Ampere(CCL/CCM/CCH)]
[Unit=OHM(CRL/CRM/CRH)]
[Unit=Volt(CVL/CVM/CVH)]
[Unit=Watt(CPL/CPM/CPH)]
Example: L “1.0,2.5,2.5,10,2.0,5.0,5.0,5.0,0,0”
SRA
Type: All Channels
Description: This command resets the Von control signal to initial state for the
ten channels in one frame. The frame will ignore the setting if the
channel does not exist.
Setting Syntax: SRA<space>“n,n,n,n,n,n,n,n,n,n”
Setting Parameters: <NR1>, 1: RESET, Other Value: no action
Example: SRA “0,0,1,1,1,1,1,0,0,0”