GEN1000 Operator Manual
6 July 22, 2019 | 287960-ENG | R00
2.6 Levelers
The unit is supplied with levelers to stabilize the chamber in it’s final location and compensate
for variations in the finished floor.
2.7 Central Management System (Optional)
For use in conjunction with the 6000 series controllers, the Conviron Central Management™
(CM) system provides a comprehensive suite of time-saving, value-added features for remote
control and monitoring of chambers, such as:
A supervisory dashboard that provides a quick overview of all operational chambers.
Thumbnail properties that show essential chamber management details.
Triggers that protect the experiment by alerting designated personnel.
Risk management that includes auto backup and restore, system protection, disaster
recovery, and file restoration.
Data management capabilities that include collection and storage.
Refer to the Conviron Central Management Operator`s Manual for complete instructions.
2.8 CMP-Link (Optional)
The CMP-LINK feature enables Argus Titan 900 full interaction of any Conviron chamber or
room that is equipped with CMP6060 control system. With CMP-LINK enabled, features of Titan
900 such as scheduling and programming, table views for data, comprehensive graphing, real-
time status (including alarms, sensors and IOs), retrieval and exporting of chamber data, time
synchronization and a remote interface are available for the chamber.
2.9 Central Alarm Contact
The Central Alarm Contact (CAC) (Figure 3-49) consists of a normally closed dry contact
energized by a control system alarm output that is connected to a Building Management System
(BMS) or an optional auto dialer. When an alarm condition occurs, that contact opens,
interrupting the circuit from the BMS (or other system) to indicate an alarm. The CAC is only
triggered by shut-down alarms.
The connection point to the BMS is a receptacle on the back panel of the chamber. This
receptacle is connected to a relay that operates when a chamber alarms. The wire connections
of the receptacle must be completed according to the electrical drawings. The electrical load
must be within the rating of the technical specifications. The receptacle is mated with the plug
and then screwed tightly together.