6.1 Disconnect the Power Before Maintaining or Servicing the Chamber
To maintain or service this chamber, it is necessary to disconnect it from power.
1. Turn the chamber power off by moving the toggle switch on the front panel to the
OFF position, as indicated by the symbol.
2. Press and hold the UPS button until the power shuts off. Verify that all three LED
indicators are off.
3. Turn off and lock the building disconnect switch provided during the installation.
4. Remove the load plug from the UPS.
5. Verify the power is off between contacts L1A and NA after opening a service access
panel as an additional precaution.
6.2 Clean the Chamber
For interior cleaning, dampen a clean towel or rag outside the unit, and carefully wipe the unit
down. Do not use abrasive cleaners. Mild detergents solutions are suitable for most cleaning
requirements.
Use glass cleaner on both the interior and the exterior of the glass viewport window, if present.
6.3 Replace Fluorescent or LED Lamps
Inspect the lamps daily to ensure that all lamps are functioning properly and replace poorly lit or
flickering lamps to ensure unit performance. The frequency of lamp changes will be determined
by application.
Warning: Burn hazard
Personal injury could result from contacting hot surfaces within the chamber.
The refrigeration system components and heater element become hot during normal
operation.
Allow the hot refrigeration components and heater elements to cool to the touch
before service.
Warning: Potential user injury after service
Personal injury to the users could result from not replacing the access panels after
service.
Ensure the access panels are replaced and secure before operating the chamber
after service.
Warning: Burn hazard
Do not touch the ends of the lamp tubes. Fluorescent lamps operate at high
temperatures and present a burn hazard.