July 22, 2019 | 287960-ENG | R00 55
8 TECHNICAL SPECIFICATIONS
Conviron maintains a policy of continual improvement and reserves the right to change the
technical characteristics of the GEN1000 without prior notice.
Specifications
GEN1000
Short Plant
(SH)
Tissue Culture
(TC)
Tall Plant
(TA)
Incubator
(IN)
Crated Weight (lbs / kg)
700 / 318
Exterior Dimensions
Height (in. / mm)
77 / 1956
Width (in. / mm)
41 / 1040
Depth (in. / mm)
32.5 / 825
Interior Dimensions
Height (in. / mm)
52.5 / 1330
Width (in. / mm)
37.5 / 953
Depth (in. / mm)
24.3 / 617
Volume (ft
3
/ L)
27.6 / 78
Power Requirements
The chamber will tolerate ±10% voltage fluctuation from the rated voltage on the serial plate.
A voltage stabilizer must be used if the fluctuation is greater than ±10%.
60 Hz
120Vac, 1Ph, N, PE, 60Hz, 20A
60 Hz dehumidification
120Vac, 1Ph, N, PE, 60Hz, 30A
50 Hz
230Vac, 1Ph, N, PE, 50Hz, 16A
Customer Alarm Contact
230Vac, 0.5A maximum
Environmental Requirements
Temperature
95°F (35°C) maximum
Humidity
Up to 55% RH, non-condensing