Operation Zeroing the CubiScan 125
CubiScan 125 31
Zeroing the CubiScan 125
Tap the [Zero] button on the touchscreen to “zero” the CubiScan 125 (set all
empty measurements and weight to zero). The weight when the platform is
empty must be set to zero for the CubiScan 125 to operate properly. The
CubiScan 125 tries to zero itself automatically. However, you may need to
use this option in the following circumstances.
If, during a long measuring session, environmental conditions
(temperature and humidity) have changed noticeably.
If you suspect that the last zeroing was in error (something was on the
platform).
NOTE
Make certain that the platform is free of all objects before using Zero. If not, the
zero reading will not be accurate.