ChemLogic® CL96 Operating Manual
6.2.C Setup -> Outputs
When the Point Setup button is touched the Outputs, Sub-Menu appears as described below. The outputs
screen allows calibration & viewing of all output modules installed on the CL96.
6.2.c.1 Setup -> Outputs -> Alarms
When the Test Faults button is touched the screen below is displayed. This same screen is used to test the
24V output models and/or the relay contact if installed. For each installed analyzer, the touch button allows the
operator to toggle the outputs active/inactive. Depending on the system setting for Energize/De-energized
relays the output is set accordingly. On the left side of each analyzer tray display is a box which is colored
appropriately for the selected faults. Below that is another box that shows which outputs are ‘active’.
In figure 6.2.c.1 below both the critical fault for points 1-16 and general fault on points 17-32 are selected.
Since the critical fault is higher priority the color of the box is orange (see table 5.2). If energized relays have
previously been setup on the unit then the critical fault relay for analyzer A1 would not be off while the Alarm
Level 1, Alarm Level 2, and general fault relays for analyzer A1 would be on.
Figure 6.2.c.1