ChemLogic® CL96 Operating Manual
7.6 End of line particular filters
End of line (point detection particulate filters which protect the CL96 from damage are required on all points
including points not being monitored. Table 7.1 details the type of filter required for each gas. Filters must be
replaced on a regular basis as shown in the table. Filter orientation is not critical in either application.
Corrosive Use Non-Corrosive Use
Part Numbers:
Blue Housing: 60009
Filter Membrane: 60010
Part Number
2-800-013
Part Number:
780248
Mineral Acids
Chlorine
Oxidizers (Excluding NO
2
)
Ammonia/Amines
SO
2
Hydrides
Phosgene
NO
2
Table 7.1
IMPORTANT : All points require filtration to prevent dust accumulation in tubing and internal
damage to the CL96. Dust that collects in the tubing or the internal system may cause sample
loss and inaccurate gas concentration readings.