ChemLogic® CL96 Operating Manual
6.3.e.3 Parallel Sampling
The CL96 Parallel Sampling option allows a sample line to be shared between 2 or 3 analyzers, allowing
multiple gases to be analyzed from a single sample line. The main sample line requires ¼”OD x 3/16” ID FEP
Teflon tubing connected by ¼” T and elbow fittings (Teflon, PVDF or Polypropylene only) to the other analyzers.
Each analyzer that is used will increase the flow on the sample line resulting in reduced transfer times while
the individual flow to each point on the manifolds will remain the same as the individual points not using the
parallel sampling. You must enable parallel sampling in the CL96 Configuration settings and check parallel
sampling for each point in the individual point set up screen.
See Figure A for an example of Parallel Sampling set up.
Figure A