ChemLogic® CL96 Operating Manual
7.4 ChemLogic
®
Cassette
The ChemLogic cassette paper tape has an expiration date printed on the label. Expired tape should be
disposed of and replaced with new tape to assure proper gas concentration readings. Each DOD ChemLogic®
tape cartridge for the CL96 will last for 120 days under normal usage. See Appendix A for ordering information.
7.5 ChemLogic Cassette Installation
1. From the main menu touch the ‘Load Tape’ button (section xxx)
2. Open the maintenance door. (section 7.1)
3. On the screen touch the ‘OPEN/CLOSE GATE’ button to open up the gate
4. (For 32-point systems only) Remove the upper take-up real by gently pulling.
5. Remove the lower take-up reel by gently pulling and dispose properly.
6. (For 32-point systems only) Remove the upper empty tape reel.
7. Remove the lower empty tape reel and install as the new lower take-up reel.
8. (for 32-point systems only) Install the empty reel as the new upper take up reel.
9. Secure the new lower ChemLogic cassette tape on the bottom take up reel. The tape should be around
the bottom in a clockwise direction as shown.
10. Feed the tape as shown with the arrows in figure 7.1.
a. From the lower tape reel
b. Around the bottom tape guide
c. Through the opening between the right-side optic blocks
d. Between the middle tape guides
e. Through the opening between the left side optic blocks
f. Between the rubber roller and the capstan
g. Around the tape guide on to the take-up reel
h. Fold the end of the tape and insert into the slot in the empty tape reel. Be sure the tape is wound
clockwise around the take up reel.
i. Turn the take-up at least 2 full turns to secure the tape.
j. (for 32-point systems only) Repeat these steps for the upper tape.
11. On the screen touch the ‘OPEN/CLOSE GATE’ button to close the gate
12. On the screen touch the ‘Reset Counter’ button. (See section 6.2.c)
13. On the screen touch the ‘Verify Optics’ button to calibrate the optics systems.
WARNING : Keep fingers clear of the gate system while touching the open/close gate
button.