ChemLogic® CL96 Operating Manual
6.3.e.3.1 Parallel Sampling Transport Times
The table below shows the time required for a sample to travel from the sampling location or the end of the
sample line to the CL96. The transport times are based on various tubing lengths and sample lines.
6.3.e.4 Reference Validation Fault
CL96 software version 16.0620 (2016), implemented three features to take action when a reference
validation message occurs. These features can be enabled or disabled in the FactorySettings screen.
To enable, check the desired box and save the configuration. To disable uncheck the box and save the
configuration.
The CL96 advanced reference diode allows the optic system to verify the proper intensity of the LED’s
continuously (See Technical Note - ChemLogic Optic Block Diagram). In the rare event a fluctuation of an LED
occurs, the reference diode catches the fluctuation and previously would report an informational message
being logged for the specific analyzer and point combination in the event log with no faults occurring. Software
version 16.0620 provides three optional actions or faults when a reference optic validation occurs.
1. REFERENCE FAULT CRITICAL
a. When a reference validation occurs, a critical fault is triggered and analysis mode is exited for the
specific analyzer (16 points)
b.
2. REF FAULT DISABLE BLOCK