ChemLogic® CL96 Operating Manual
• Attach the LineBlack wire to the Power Filter at the label for Screw #1, tighten securely.
• Push the plastic cover back into place on the Power Filter over screws connections #1 and #2.
• Locate the Protective Earth (PE) Grown stud as shown in figure G.2.4.
Figure G.2.4
• Remove the top nut from the PE ground stud
• Leave the star washer on the bottom of the ground stud touching the cabinet, then place the ring
terminal from the green wire over the star washer.
• Replace the nut on top of the ring terminal and tighten securely.
• NOTE: Do not attach any wires other than the Green ground wire from the A/C power cord to the PE
Ground terminal.
G.3 Connecting to the CL96
Warning
: Do NOT use the A/C power opening to bring any
other wires into the secondary enclosure. Analog wiring and
the ethernet cable must be brought in through a separate
opening dependent on the user’s design.
Drill a second hole in the side of the external cabinet at the desired location. Bring the shielded ethernet cable
into the cabinet through the hole and connect as described below.
G.3.1 Connecting to the CL96 with Ethernet Cable
All secondary enclosures connect to the CL96 with a shielded RJ45 cable (sold separately).
• Attach one end of the cable to the CL96. Depending on the configuration of your machine the
connection point will vary. Contact DOD if you are not sure where to plug the ethernet cable.
• Attach the other end of the cable to the top connector on the EK1101 module as shown in Figure G6.