ChemLogic® CL96 Operating Manual
4. From the FACTORY Menu SETTINGS
6.3.e.1 Minimum Visible Stain
The Minimum Visible Stain (MVS) feature was implemented as an enhancement (in 2017) as part of the
software version 17.0130 update. MVS allows color change or the stain to continue to accumulate during the
same window, or on the same spot of the ChemLogic cassette, until the minimum count threshold is achieved.
The default setting for MVS is 250 counts. MVS is user adjustable from 0 to 400 counts. 0 counts would
disable the MVS feature. When enabled, the MVS software will not trigger a tape advance during a gas event
until the set MVS value is reached. MVS set at 250 counts and above should result in the stain being visible to
the human eye during a gas event.