ChemLogic® CL96 Operating Manual
The typical counts required to see a stain with the human eye is approximately 200 counts. This varies slightly
by gas and ChemLogic cassette.
The MVS software feature can be accessed and modified by following the below path.
From the FACTORY Menu SETTINGS
MVS was implemented to allow a stain or color change continue to accumulate to pre-set count changes. The
purpose is to allow low level concentrations to accumulate on the same spot making the stain visible to the
human eye. Previous software versions advanced tape when set count changes were reached. During low level
detection this could result in tape advances and low-level readings without visible stains. MVS reduces the
possibility of low-level detection without visible stains.