ChemLogic® CL96 Operating Manual
Chapter 7 Maintenance and Disposal
DANGER: Disconnect power before servicing
7.1 Returning to a Safe State (Post-Service)
Before returning the CL96 to service after maintenance or perform verify the following checks:
Verify all A/C power connections are secured properly.
Check all ground wire connections are secured properly to each panel on the unit.
Verify each analyzer tray is installed on the rails and connected properly.
Check tubing connections on both sides of each pump.
Verify all sample tubing and exhaust tubing connections on the unit.
7.2 Maintenance Door Access
The maintenance door is used to access the ChemLogic cassette. Two keyed latches secure the door. Turn
both latches with the keys to open the maintenance door and be sure to secure both latches when closing the
door.
7.3 Control Panel Door
WARNING : Turn off and remove power from the system prior to opening the control panel
door.
To open the control panel door:
Insert the key provided into the slot and rotate counterclockwise to unlock the door.
Turn the latch counterclockwise to unlatch the door.
Life the control panel access door until it locks (listen for click)
The door should remain in the up position until the lock is released.
When service is completed be sure to close the service door and secure the keyed latch to the closed position.
Verify that the service door cannot be pulled open secure the door using the key to lock the door.
IMPORTANT : The maintenance door should remain closed and latched except when servicing
the ChemLogic® cassette.
IMPORTANT : The control panel access door should remain closed and latched except when
servicing the system.