ChemLogic® CL96 Operating Manual
6.3.e.2 Dual Pump Option
The CL96 Continuous Gas Monitor has an optional dual pump feature. This option includes an additional pump
for each 16-point analyzer, resulting in up to 80% reduced transport time. This dual pump feature also acts as a
backup pump in the event of a pump failure. For each 16-point set, two pumps are installed (see Figure 1).
Check valves are connecting the two pumps in each 16-point analyzer (see Figure 2). In the event, one pump
fails the other pump will take over for the failed pump. Flow rates will drop from the data provided in Table 2 to
the data provided in Table 1, but all points will remain in analysis. A general fault will be issued stating that the
pump has failed.
Below includes the data from the current standard CL96 transport times and flow rates (1) and the dual pump
option transport times and flow rates (2).
1. Single pump transport times and flow rates – STANDARD CL96 CONFIGURATION
Table 1